Size specific dose estimate (SSDE) for estimating patient dose from CT used in myocardial perfusion SPECT/CT

Document Type : Original Article

Authors

Department of Nuclear Medicine, Jawaharlal Institute of Postgraduate Medical Education and Research, Puducherry, India

Abstract

Objective(s): Size specific dose estimate (SSDE) is a new parameter that includes patient size factor in its calculation. Recent studies have produced mixed results on the utility of SSDE, especially when automatic exposure control (AEC) was used. The objective of the study was to find out if there is a relationship between patient size and each of the parameters, SSDE and CTDIvol, when AEC is used.
Methods: CT data of consecutively selected 111 patients were included for analysis. CTDIvol values of the CT scans were extracted for each patient. Effective diameter of each patient was calculated as geometric mean of anteroposterior and lateral diameters measured on axial CT images. Corresponding conversion factors for effective diameters were obtained from American Association of Physicists in Medicine (AAPM) report 204. SSDE was obtained as the product of CTDIvol and conversion factor values. Linear regression model was used to evaluate the relationship between patient size and the parameters SSDE and CTDIvol.
Results: Mean weight was 62 (11.5) and range was 34 - 103 kg. Median CTDIvol (mGy) on AEC mode was 7.27(IQ range 7.27, 7.65) and mean effective diameter was 26.2 cm (2.4). Mean SSDE (mGy) was 10.6 (0.84). Good positive correlation was obtained between CTDIvol and effective diameter (r=0.536; p<0.0005). Strong inverse correlation was noted between SSDE and effective diameter (r=-0.777; p<0.0005). Linear regression model for establishing relationship between CTDIvol and effective diameter showed slope of 0.314mGy/cm (R=0.561; R2=0.314; P<0.0005) whereas between effective diameter and SSDE slope was -0.23mGy/cm (R=0.676; R2=0.457; P< 0.0005).
Conclusion: The study shows that CTDIvol and SSDE vary but divergently, with patient size. SSDE is a better estimate of patient radiation dose from CT of MPI SPECT/CT than CTDIvol in systems that use automated exposure control.

Keywords

Main Subjects


1. CDC. ALARA - As Low as Reasonably Achievable [Internet]. Centers for Disease Control and Prevention. 2015 [cited 2019 Apr 29]. Available from: https://www.cdc.gov/ nceh/radiation/alara.html.
2. Bauhs JA, Vrieze TJ, Primak AN, Bruesewitz MR, McCollough CH. CT Dosimetry: Comparison of Measurement Techniques and Devices. RadioGraphics. 2008; 28(1):245–53.
3. Schindera ST, Nelson RC, Toth TL, Nguyen GT, Toncheva GI, DeLong DM, et al. Effect of Patient Size on Radiation Dose for Abdominal MDCT with Automatic Tube Current Modulation: Phantom Study. Am J Roentgenol. 2008; 190(2):W100–5.
4. Miyazaki O, Horiuchi T, Masaki H, Nosaka S, Miyasaka M, Tsutsumi Y, et al. Estimation of adaptive computed tomography dose index based on body weight in pediatric patients. Radiat Med. 2008; 26(2):98–103.
5. McCollough CH, Leng S, Yu L, Cody DD, Boone JM, McNitt-Gray MF. CT Dose Index and Patient Dose: They Are Not the Same Thing. Radiology. 2011; 259(2):311–6.
6. AAPM Reports - Size-Specific Dose Estimates (SSDE) in Pediatric and Adult Body CT Examinations [Internet]. [Cited 2018 Sep 4]. Available from: https://www.aapm.org/pubs/reports/ detail.  asp?  docid = 143
7. Brady SL, Kaufman RA. Investigation of American Association of Physicists in Medicine Report 204 size-specific dose estimates for pediatric CT implementation. Radiology. 2012; 265(3):832–840.
8. Bhatt S, Srivastava AK, Meena N, Thakur S. Appraisal of radiation dose with 64-slice computed tomography perfusion in lung cancer patients with special reference to SSDE: An initial experience in a tertiary care hospital. Indian J Radiol Imaging. 2017; 27(4):389–96.
9. Christner JA, Braun NN, Jacobsen MC, Carter RE, Kofler JM, McCollough CH. Size-specific Dose Estimates for Adult Patients at CT of the Torso. Radiology. 2012 1; 265(3):841–7.
10. Tsujiguchi T, Obara H, Ono S, Saito Y, Kashiwakura I.  Consideration of the usefulness of a size-specific dose estimate in pediatric CT examination. J Radiat Res (Tokyo).  2018 1; 59(4): 430–5.
11. Strauss KJ. Dose indices: everybody wants a number. Pediatr Radiol. 2014 Oct; 44 Suppl 3:450–9.
12. Nickoloff EL, Dutta AK, Lu ZF. Influence of phantom diameter, kVp and scan mode upon computed tomography dose index. Med Phys. 2003; 30(3):395–402.
13. Abdollahi H, Shiri I, Salimi Y, Sarebani M, Mehdinia R, Deevband MR, et al. Radiation dose in cardiac SPECT/CT: An estimation of SSDE and effective dose. Eur J Radiol. 2016; 85(12):2257–61.