Differential impact of multi-focus fan beam collimation with L-mode and conventional systems on the accuracy of myocardial perfusion imaging: Quantitative evaluation using phantoms

Document Type : Original Article

Authors

1 Program in Health and Welfare, Graduate School of Comprehensive Scientific Research, Prefectural University of Hiroshima

2 Department of Radiology, Kurashiki Central Hospital

3 Program in Biological System Sciences, Graduate on School of Comprehensive Scientific Research, Prefectural University of Hiroshima

4 Department of Nursing,Hyogo University of Health Sciences

Abstract

Objective(s): A novel IQ-SPECTTM method has become widely used in clinical studies. The present study compares the quality of myocardial perfusion images (MPI) acquired using the IQ-SPECTTM (IQ-mode),conventional (180° apart: C-mode) and L-mode (90° apart: L-mode) systems. We assessed spatial resolution, image reproducibility and quantifiability using various physical phantoms.
Methods: SPECT images were acquired using a dual-headed gamma camera with C-mode, L-mode, and IQ-mode acquisition systems from line source, pai and cardiac phantoms containing solutions of 99mTc. The line source phantom was placed in the center of the orbit and at ± 4.0, ± 8.0, ± 12.0, ± 16.0 and ± 20.0 cm off center. We examined quantifiability using the pai phantom comprising six chambers containing 0.0, 0.016, 0.03, 0.045, 0.062, and 0.074 MBq/mLof 99m-Tc and cross-calibrating the SPECT counts. Image resolution and reproducibility were quantified as myocardial wall thickness (MWT) and %uptake using polar maps.
Results: The full width at half maximum (FWHM) of the IQ-mode in the center was increased by 11% as compared with C-mode, and FWHM in the periphery was increased 41% compared with FWHM at the center. Calibrated SPECT counts were essentially the same when quantified using IQ-and C-modes. IQ-SPECT images of MWT were significantly improved (P<0.001) over L-mode, and C-mode SPECT imaging with IQ-mode became increasingly inhomogeneous, both visually and quantitatively (C-mode vs. L-mode, ns; C-mode vs. IQ-mode, P<0.05).
Conclusion: Myocardial perfusion images acquired by IQ-SPECT were comparable to those acquired by conventional and L-mode SPECT, but with significantly improved resolution and quality. Our results suggest that IQ-SPECT is the optimal technology for myocardial perfusion SPECT imaging.

Keywords

Main Subjects


  1. Jinghan Y, Xiyun S, Zuo Z, Da Silva. Iterative SPECT Reconstruction Using Matched Filtering for Improved Image Quality. Nuclear Science Symposium conference IEEE. 2006;2285-7.
  2. Hughes T, Scherbinin S, Celler A. A multi-center phantom study comparing image resolution from three state-of art SPECT-CT systems. J Nucl Cardiol. 2009; 16:914-26.
  3. Onishi H, Motomura N, Fujino K, Natsume T, Haramoto Y. Quantitative performance of advanced resolution recovery strategies on SPECT images: Evaluation with use of digital phantom models. Radio Phys Technol. 2013; 6:42-53.
  4. Myers RW, Haseman MK, Durbin M. Halftime acquisition of MPI with maintained image quality. J Nucl Cardiol. 2006; 13:S14-5.
  5. DePuey EG, Gadiraju R, Clark J, Thompson L, Ansetett F, Shwartz SC. Ordered subset expectation maximization and wide beam reconstruction "half-time" gated myocardial perfusion SPECT functional imaging: a comparison to "full-time" filtered backprojection. J Nucl Cardiol 2008; 15:547-63.
  6. Ali I, Ruddy TD, Almgrahi A, Anstett FG, Wells RG. Half-time SPECT myocardial perfusion imaging with attenuation correction. J Nucl Med. 2009; 50:544-62.
  7. Lewin HC, Hyun MC. A clinical comparison of UN upright triple-head digital detector system to a standard supine deal-head gamma camera [abstract]. J Nucl Cardiol 2005, 12:113.
  8. Funk T, Kirch DL, Koss JE, Botvinick E, Hasegawa BH. A novel approach to multipinhole SPECT for myocardial perfusion imaging. J Nucl Med 2006;47: 595-602.
  9. Vija H, Chapman J, Ray M. IQSPECT technology White Paper. Siemens Medical Solutions USA Molecular Imaging; 2008:1-7.
  10. Sharir T, Ben-Haim S, Merzon K, Prochorov V, Dickman D, Ben-Haim S, et al. High-speed myocardial perfusion imaging: Initial clinical comparison with Conventional dual detector anger camera imaging. J Am Coll Cardiol Imaging 2008; 1:156-63.
  11. Vija H, Malmin R, Yahil A, Zeintl J, Bhattcharya M, Rempel TD, et al. A method for improving the efficiency of myocardial perfusion imaging using Conventional SPECT and SPECT/CT imaging systems. Paper represented at IEEE Nuclear Science Symposium conference 2010; 6:3433-7.
  12. Hawman PC, Haines RJ. The cardiofocal collimator: A variable focus collimator for cardiac SPECT. Phys Med Biol 1994; 39:439-50.
  13. Corbett J, Meden J, Ficaro E. Clinical validation of attenuation corrected cardiac imaging with IQ-SPECT SPECT/CT [abstract]. J Nucl Med. 2010; 51(suppl 2): 1722.
  14. Knoll P, Kotalova D, Köchle G, Kuzelka I, Minear G, Mirzaei S, et al. Comparison of advanced iterative reconstruction methods for SPECT/CT. Z Med Phys. 2012; 22:58-69.
  15. Cerqueira MD, Weissman NJ, Dilsizian V, Jacobs AK, Kaul S, Laskey WK, et al. Standardized myocardial segmentation and nomenclature for tomographic imaging of the heart. Circulation 2002;105:539-42.
  16. Paramithas A, Irwin AG. Characterization of the IQSPECTTM cardiac SPECT system. EANM2011 [abstract]. Eur J Nucl Med Mol Imaging 2011; 38 (suppl 2):103P.
  17. Imbert L, Poussier S, Franken PR, Songy B, Verger A, Morel O, et al. Compared performance of high-sensitivity camera dedicated to myocardial perfusion SPECT: a comprehensive analysis of phantom and human images. J Nucl Med 2012; 53:1897-1903.
  18. Zeintl J, Vija AH, Yahil A, Hornegger J, Kuwert T. Quantitative accuracy of clinical 99mTc SPECT/CT using ordered-subset expectation maximization with 3-dimensional resolution recovery, attenuation, and scatter correction. J Nucl Med 2010; 51: 921-8.
  19. Liu YH, Lam PT, Sinusass AJ, Wackers FJ. Differential effect of 180 degrees and 360 degrees acquisition orbits on the accuracy of SPECT imaging: Quantitative evaluation in phantoms. J Nucl Med 2002; 43: 1115-24.
  20. Goetze S, Brown TL, Lavely WC, Zhang Z, Bengel FM. Attenuation correction in myocardial perfusion SPECT/CT: Effects of misregistration and value of reregistration. J Nucl Med 2007; 48:1090-5.